HeartKit Rhythm Tutorial

This tutorial shows running an end-to-end heart rhythm classifier on the Apollo 4 EVB. The basic flow chart is depicted below.

flowchart LR
    S[1. Collect] -->| | P[2. Preprocess] --> M[3. Rhythm Model] --> L[4. Display]

In the first stage, 4 seconds of sensor data is collected- either directly from the MAX86150 sensor or test data from the PC. In stage 2, the data is preprocessed by bandpass filtering and standardizing. The data is then fed into the CNN network to perform inference. Finally, in stage 4, the ECG data will be classified as normal (NSR), rhythm (AFIB/AFL) or inconclusive. Inconclusive is assigned when the prediction confidence is less than a pre-defined threshold (e.g. 90%).

Note

A reference rhythm model (./evb/src/arrhythmia_model_buffer.h) is included and can be used to quickly evaluate the hardware. The model is trained on Icentia11k dataset that has the associated non-commercial license. The model is intended for evaluation purposes only and cannot be used for commercial use without permission.

Demo Setup

Please follow EVB Setup Guide to prepare EVB and connect to PC. To use the pre-trained model, please skip to Run Demo Section.

1. Train the rhythm model w/ given configuration

heartkit \
    --task rhythm \
    --mode train \
    --config ./configs/train-rhythm-model.json

Note

Due to the large dataset and class imbalance, the batch size and buffer size are large for rhythm training to ensure properly shuffling of patients as well as classes. The first epoch will take much longer as it fills up this buffer. To train on dedicated GPUs, it's recommended to have at least 10 GB of VRAM.

2. Evaluate the rhythm model's performance

heartkit \
    --task rhythm \
    --mode evaluate \
    --config ./configs/evaluate-rhythm-model.json

3. Export the model into a TFLM header file

heartkit \
    --task rhythm \
    --mode export \
    --config ./configs/export-rhythm-model.json

The exported model will be placed into ./evb/src/arrhythmia_model_buffer.h. Please review ./evb/src/constants.h and ensure settings match configuration file.

Run Demo

Please open three terminals to ease running the demo. We shall refer to these as EVB Terminal, REST Terminal and PC Terminal.

1. Run client on EVB

Run the following commands in the EVB Terminal. This will compile the EVB binary and flash it to the EVB. The binary will be located in ./evb/build.

make -C ./evb
make -C ./evb deploy
make -C ./evb view

Now press the reset button on the EVB. This will allow SWO output to be captured.

2. Run REST server on host PC

In REST Terminal, start the REST server on the PC.

python -m heartkit.demo.server

3. Run client and UI on host PC:

In PC Terminal, start the PC client (console UI).

heartkit \
    --task rhythm \
    --mode demo \
    --config ./configs/rhythm-demo.json

Upon start, the client will scan and connect to the EVB serial port. If no port is detected after 30 seconds, the client will exit. If successful, the client should discover the USB port and start updating UI.

4. Trigger start

Now that the three tasks are running, press either Button 1 (BTN1) or Button 2 (BTN2) on the EVB to start the demo. Pressing Button 1 will use live sensor data whereas Button 2 will use test dataset supplied by the PC. In EVB Terminal, the EVB should be printing the stage it's in (e.g INFERENCE STAGE) and any results. In PC Terminal, the PC should be plotting the data along with classification results. Once finished, Button 1 or Button 2 can be pressed to stop capturing.

evb-demo-plot

Tip

Please use a monospaced font in the terminal for proper alignment of the plot.

To shutdown the PC client, a keyboard interrupt can be used (e.g [CTRL]+C) in PC Terminal. Likewise, a keyboard interrupt can be used (e.g [CTRL]+C) to stop the PC REST server in REST Terminal.